Model:JXDC-96
Brand:Jingxin
Specifications :Set
Introduction:Nitrogen purging instrument is a high-precision temperature control instrument formed by the combination of microcomputer processing and PID control.
1. Product introduction
The nitrogen purging instrument is a high-precision temperature control instrument formed by the combination of microcomputer processing and PID control. Its working principle is to blow nitrogen quickly, continuously and controllably to the surface of the heated sample, according to the evaporation speed and boiling point of the concentrated solvent , Set the heating temperature to achieve rapid concentration of a large number of samples. This instrument uses an aluminum plate heating module, which has good heat transfer and uniform heat transfer, which is conducive to rapid heating and rapid temperature control. Blow nitrogen to the surface of the sample to achieve oxygen-free concentration of the liquid sample. The blowpipes are independent of each other and will not cause cross-contamination. The air needle channels on the air distribution assembly can be used in combination or individually. The system has high temperature control accuracy, wide temperature control range, temperature numerical control digital display, and convenient temperature calibration. The product has beautiful appearance, simple operation, safe and reliable use.
2. Field of Use
The main application areas of this product:
★ Pesticide residue analysis: such as vegetables, fruits, grains, plant tissues
★ Environmental analysis: such as drinking water, groundwater and polluted water samples
★ Biological analysis: such as sample preparation in hormone analysis, liquid phase, gas phase and mass spectrometry analysis
★ Food and beverages: such as milk, wine, beer, etc. ★ Pharmaceutical drug testing: such as traditional Chinese medicine, drug screening
Product parameters:
Main parameters | Parameter range |
Model | JXDC-96 |
Temperature control range | Room temperature+5°C -160°C |
Temperature control accuracy | ±0.5℃(@40℃) |
Temperature control accuracy | ±1℃(@120℃) |
Temperature uniformity | ±0.5℃ |
Display accuracy | 0.1℃ |
Timing range | 1-99h59min/∞ |
Heating time | ≤15 minutes (25°C to 160°C) |
Maximum lifting stroke | 150mm |
Maximum gas flow | 15L/min |
Maximum gas pressure | 0.05Mpa |
Number of modules | 1 |
Module capacity | See product accessories for details |
Maximum power | 500W |
Voltage specifications | 220V 50/60Hz |
Size | 220*310*385mm |
Weight | 7.5Kg |
Product acessories:
Module model | Adapter board specification | Module size (mm) |
MD17 | 96-well 0.2mL PCR plate | 153X95.5X60 |
MD18 | 96-well microtiter plate (flat bottom) | 153X95.5X60 |